|
How to Select Pressure TransducersPressure Range
AccuracyApplication Considerations
Pressure RangeWhat will be the pressure range? A good rule of thumb: figure out what your anticipated pressure spikes will be and then pick a transducer rated 25% higher than your highest spike. An additional margin is suggested where "high cycling" may occur. Consult with Trans-Metrics if you are unsure how to select the most accurate model with appropriate proof-pressure and burst-pressure protection. What will be the units of
measurement?
For example: What is the pressure reference?
Local, absolute, sealed or differential? All transducers measure pressure relative to a second or reference pressure. The vast majority of pressure measurements are made relative to the local atmospheric pressure. In the USA, these are called gage type transducers (psig). When a transducer measures pressures relative to the local atmospheric pressure and less than the ambient pressure, it is called a vacuum type transducer (psiv). A transducer that measures both positive and negative pressures is called a compound transducer (psig). A sealed gage transducer (psis) measures pressure relative to the standard atmospheric pressure at sea level (14.69 psia). In moist or humid environments, sealed gage transducers are frequently used to protect the sensors internal components. An absolute transducer (psia) measures pressure relative to a vacuum (zero pressure). A differential transducer (psid) has an output that is the difference between the pressures applied to its two pressure ports. AccuracyWhat are the components of
transducer accuracy? While the terms "accuracy" and "static error band" are frequently interchanged, they are not the same. Many factors can contribute to a transducers accuracy. Actually, the term "error" is preferred for specifications and other specific descriptions of transducer performance. The static error band is, however, a good measure of the accuracy that can be expected at constant temperature. The static error band consists of the three following components: Non-linearity is the deviation of a calibration curve from a specified straight line. The best straight line (BSL) is a line between two parallel lines that enclose all output vs. pressure values on the calibration curve. Hysteresis is the maximum difference in output at any pressure within the specified range, when the value is first approached with increasing and then with decreasing pressure. Non-repeatability is the inability of a transducer to reproduce output readings when the same pressure is applied to the transducer repeatedly, under the same conditions and in the same direction. In addition to the three components of static error band, you may need to consider the effects of temperature. Thermal effects typically create the largest errors in pressure measurement. Thermal errors are commonly expressed in percent per degree, although this can be deceptive. An apparently very small percentage error per degree may actually be a significant problem when checked over the entire expected range. To combat thermal effects, transducers should be 100% tested and compensated over a specified temperature range (as are all of Trans-Metrics transducers). The compensated temperature range is usually smaller than the operating temperature range. Note that the thermal sensitivity shift is usually expressed as a percent of output. What output signal is needed? Low-level transducers usually have only passive devices (strain gages and resistors) in their electrical circuits. Thus they are typically more rugged and are able to operate over a wider temperature range. Signal conditioning is external. The highest accuracy over the broadest environmental conditions can be provided with low-level transducers. A typical low-level output is 3 mV/V. High-level transducers have internal electronic circuits to condition the excitation or supply voltages and the signal. They are easy to install because most of the signal conditioning is already provided. Some typical high-level output are 0-5 VDC, 0-10 VDC, 1-5VDC, and 4-20 mA. NOTE: Most high-level Trans-Metrics transducers have zero and FSO (span) adjustment pots; therefore, they are field-adjustable. What will be the output at zero
pressure? Trans Metrics offers all of the above and more. What degree of transducer
interchangeability is required? What power is available? In remote locations where power is supplied by battery or solar power, a low-consumption transducer is needed. Circuits would typically consist of a low dropout regulator, a voltage doubler, a negative voltage converter and a signal amplifier. Trans-Metrics P100 Series with L circuit type is an example of a low-power consumption transducer series. With excitation voltages of 4-26 VDC, it can provide a regulated 0-5 VDC output. If the application's power supply is unregulated, the transducer package needs internal voltage regulation. All high-level Trans-Metrics transducers have internal voltage regulation if the power supply is unregulated. What type of electrical
connections are there for the transducers? There are many makes and models of electrical connectors suitable for transducer use. When the conductors exit directly from the transducer they may be called a cable output, pigtails or flying leads. Conduit connections may also be included with this approach. It is important that the strain relief and seal are adequate for the application regardless of the method. Note that applications with high temperatures, high humidity (including submersion), or high shock and/or vibration may require special treatment. Discuss your application with your Trans-Metrics Application Engineer. Transducer Circuit-Type: two-,
three- or four-wire? Transducers with high-level outputs may have two-, three- or four-wire connections. The 4-20 mA current loop transducer is a two-wire type where the transducer, power supply and readout are all connected in series. The three-wire circuit has a common return for both the excitation and signal. This type of circuit may have the zero balance elevated so that the signal will vary from between 1-5 VDC or 2.5-7.5 VDC (such as the Trans-Metrics P100, D circuit type). Some Trans-Metrics three-wire circuits (such as our P100 C and L circuit type) have an internal negative voltage source so that the transducer will have zero volts out at zero pressure. The four-wire circuit has a separate return connection for the excitation and signal. The signal is differential with a common mode voltage different from the excitation return. Note that on this type of transducer, the power supply and signal lines cannot be connected to the same point. What are the pressure media to
be measured? How will the transducer be
connected to the pressure source? Anther type of pressure port is the flush mount. Flush-mounting transducers, like Trans-Metric's P37 and P38 models, do not have any cavities, making them easier to clean. Please consult instruction manuals for mounting recommendations. By using adapters, virtually any transducer can be connected to almost any pressure source. What environment will the
transducer be in? Additional environmental considerations include temperature, vibration and EMI/RFI. Trans-Metrics can offer recommendations for all of these problems. Important: be sure to note any underwater applications. Special Transducer Requirements If you don't know what requirements may be considered "special," just briefly state the application on the Quote Form. For example the statement, "I have a down hole application," will prompt the transducer vendor to ask: "How big are the well holes?" or, "What size transducer diameter will fit in the well holes?" Special characteristics do not necessarily require custom transducers. If customizing is required, please note that some manufacturers welcome custom orders while others do not. Trans-Metrics, Inc. specializes in custom transducers. |
||||||||||||||||||
|